Super-systems 20PQ Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Super-systems 20PQ. Super Systems 20PQ User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Instruction Manual

170.IU0.20Q.S00Instruction ManualP/N 3109020 PQProgrammer - ControllerXkP-0-1C2.P65 10/30/01, 9:42 AM1

Page 2

6Fig. 7.B INPUTS WIRING FOR 3-WIRE TRANSMITTERShield31117PWR+OutTXGNDFig. 7.C INPUTS WIRING FOR 4-WIRE TRANSMITTERNOTES:1) Don’t run input wires toget

Page 3

96Gr. 8- Servomotor high limit positioning - [r.H04]This parameter is available only when feedback potentiometercalibration enabling [r.H01] is equal

Page 4

97Gr. 9Run time group 9 [r.Ixx]TRACKING VALUES.NOTES:1) this run time group will be shown when the instrumentoperates in programmer mode only.2) The t

Page 5 - MOUNTING REQUIREMENTS

98Gr. 9- Group 9 default data loading - [r.I03]Range: OFF = No loading dataOn = loading dataRun time group "dF" [r.Lxx]DEFAULT RUN TIME PAR

Page 6 - DIMENSIONS

99Gr. HdRun time group "Hd" [r.Mxx]HIDDEN PARAMETERS - SMART LIMITSNOTE: This group is accessible from every group (exceptionmade for P grou

Page 7 - CONNECTION DIAGRAMS

100Gr. Hd- Relative gain of the secondary outputcalculated by SMART algorithm - [r.M05]This parameter is available only when smart function isconfigur

Page 8

101EDIT.In this mode it is possible to select and/or to create a programand to visualize and/or to modify all data of a program.When the instrument is

Page 9

102HOLDThis function temporarily stops a running program by a manualaction.The HOLD mode may be activated by the RUN pushbuttonpressure or by an exter

Page 10

103BARGRAPH DESCRIPTIONTwo 33 LED’s bargraphs are available.On the right bargraph, with orange LED’s, it is possible to show:- the operative set point

Page 11

104INDICATORS°C Lit when process variable is in Centigrade degrees.°F Lit when process variable is in Fahrenheit degrees.ST Flashing when the first st

Page 12

105Flashes when alarm 2 is in alarm state and it is notacknowledged.3 Lit when OUT 3 is used as control output and it is in ONcondition or when servom

Page 13

7B) AUXILIARY INPUTFig. 8 AUXILIARY INPUT WIRINGNOTES:1) This input is not isolated from measuring input. A double orreinforced insulation between ins

Page 14

106The first FUNC pushbutton pressure modify the middle displayindication.It will show:- the time remaining to the end of the segment, if thesegment i

Page 15

107l) Push the FUNC pusbutton again; the lower display shows" Fd." followed by the OUT failure detection alarm status:OF = no alarmAL (flash

Page 16

108When device operates in programmer mode and noprogram is running (STAND-BY mode).The upper display shows the process variable and PRGindicator is l

Page 17

109It is possible to activate the O.P.O. mode by:- serial link- by keyboard- automatically at the end of a program execution.It is possible to exit fr

Page 18 - Feedback

110OUT FAILURE DETECTION FUNCTION (OFD)The device is capable to measure and display:- the current running in a load driven by a time proportionalcontr

Page 19 - 20 mA20 mA

111LAMP TESTWhen it is desired to verify the display efficiency, put theinstrument in "normal display mode" and push t and MENUpush-buttons

Page 20

112At power up the device starts as configured at “St.Fn” [C.G06]configuration parameter.NOTES:1) When auto/manual control mode is selected by logic i

Page 21

113Gr. PGENERAL NOTES ABOUT PROGRAM EDITINGThis instrument allows to create up to 90 simple programs with afree number of segments for each program.Th

Page 22 - PRELIMINARY HARDWARE SETTINGS

114Gr. PRun time group P [r.Axx]Program editingThis group is available when the instrument operates asprogrammer only.- Program selection - [r.A01]Low

Page 23 - ON DIP

115Gr. P- Event 1 enabling - [r.A04]This parameter can be modified only when the selected programis not running, it is not a linked program and output

Page 24

8Fig. 9.B - LOGIC INPUTS IN 1, 2, 3 and 4 WIRING5859IN 2IN 35657IN 160IN 4Fig. 9.C - LOGIC INPUTS IN 5, 6, 7 and 8 WIRINGNOTES:1) Do not run logic inp

Page 25

116Gr. P- Program 1 - Segment 1 - segment duration- [r.A07]This parameter can be modified only when the selected program isnot running, it is not a li

Page 26 - General notes

117Gr. PRange: from 0 to 10.NOTE: setting trK.G equal to 0, the instrument will notuse, for this segment, the tracking function.- Program 1 - Segment

Page 27

118Gr. P- Break event reset at the end of program 1- [r.A13]This parameter can be modified only when the selected programis not running, it is not a l

Page 28

119Gr. Pfrom -100.0 to 100.0 % if the devices is configured withtwo control outputs (split range excluded).- Set point at the end of program 1 - [r.A1

Page 29

120Gr. PGeneral notes about timer eventsUp to 14 timer events can be associate to each program.Timer events are composed by 5 steps each, during the f

Page 30

121SIMPLE PROGRAM MANAGEMENTHOW TO CREATE A SIMPLE PROGRAMWhen it is desired to create a new program proceed as follows:1) select the Group P;2) selec

Page 31 - CONFIGURATION MODE

122NOTES:1) When the final set point of a ramp has been modified and thenext segment is a soak, the instrument automatically modifiesthe final set poi

Page 32

1233) select the final set point of the segment x+1:4) push "RUN" + "s" pushbuttons; the display will show the finalset point of t

Page 33

124LINKED PROGRAM MANAGEMENTHOW TO CREATE A LINKED PROGRAMIt is possible to create up to 9 linked programs (L1 to L9), eachlinked program is formed by

Page 34 - CONFIGURATION GROUP 1 [C.dxx]

125HOW TO MODIFY A LINKED PROGRAMWhen it is desired to modify a linked program it is necessary tore-edit it.HOW TO DELETE AN EXISTING LINKED PROGRAMWh

Page 35

93) The minimum active period to perform this measurement isequal to 120 ms.4) The input impedance is equal to 20 W.1415LoadCurrenttransformerD) CURRE

Page 36 - -.--- = three decimal figures

126E.ñ simple program not existing (the central display willshow the number of the not existing program).This message will be shown when a simple prog

Page 37 - Read-out

127an error message will be shown (for more details aboutthis group of error messages see "How to check aprogram").During RUN mode the arrow

Page 38

128ABORTThis function allows to definitively stop the program in progress.When it is desired to ABORT the program push the RUNpusbutton and maintain t

Page 39

129FASTThis function allows to move, forward or backward, the point ofthe program in progress 60 time faster than the programmedvalue.When it is desir

Page 40

1304) When the desired point has been reached, restart theprogram.NOTES:1) When the JUMP function is activated, the central display willshow the same

Page 41

131DISPLAY FUNCTION DURING CONTROLLER MODEWhen device is powered and it is operating in automatic mode,the upper display shows the process variable an

Page 42

1320.0 % to 99.9 % or a value within the scale configured by“ñC.E.L” [C.F09] and “ñC.E.H” [C.F10] when a scaled readoutis applied to the main output.T

Page 43

133INDICATORS function during controller mode°C Lit when process variable is in Celsius degrees.°F Lit when process variable is in Fahrenheit degrees.

Page 44

134SMART FUNCTIONIt is used to automatically optimize the control action.To enable the SMART function proceed as follows:1) push the MENU push-button

Page 45

135OUTPUT POWER OFF FUNCTION during controller modeBy this mode it is possible to disable the process control.In this mode, the instrument operate as

Page 46

10E.1) RELAY OUTPUTSFig. 11.A RELAY OUTPUTS 1,2,3 and 4 WIRINGOUT 123NC24C25NOOUT 227CNO28OUT 329NO - OUT 3OUT 430C - OUT 3/431NO - OUT 426NCThe out

Page 47

136ERROR MESSAGESOVERRANGE, UNDERRANGE AND SENSOR LEADSBREAK DETECTIONON THE MAIN INPUTThe device is capable to detect a fault condition on the mainin

Page 48 - --.-- = Two decimal figures

137sensor leads break condition.When the sensor leads break condition is detected, theinstrument shows “OPEn” on the upper display.Note: On the linear

Page 49

138ON THE AUXILIARY INPUTWhen the remote set point trim signal exceeds the full scalevalue programmed by "A.In.t" [C.d10], an OVERRANGEcondi

Page 50

139Configuration and run time parameter errors.When one of this error is detected, following the proceduresdescribed for standard parameter modificati

Page 51

140Calibration parameter errorsThe instrument will show on the lower display one of thefollowing indications:ñ.I.tc = main thermocouple input calibrat

Page 52

141GENERAL INFORMATIONSGENERAL SPECIFICATIONSCase: Polycarbonate black color; self-extinguishing degree: V-0according to UL 746C.Front protection: des

Page 53

142Temperature drift: (CJ excluded)< 200 ppm/°C of span for mV and TC ranges 3, 4, 7, 18, 19, 22.< 300 ppm/°C of span for mA/V< 250 ppm/°C of

Page 54

143C) LINEAR INPUTSRead-out: keyboard programmable between -1999 and +9999.Decimal point: programmable in any positionSTANDARD RANGES TABLEAUXILIARY I

Page 55

144STANDARD RANGES TABLELOGIC INPUTSThe instrument may be supplied with 11 logic inputs program-mable as:- Program selection- RUN/HOLD mode selection-

Page 56

145Proportional Band: programmable from 0.5% to 999.0% of theinput span.Setting a PB equal to 0 the control action becomes ON/OFF.Hysteresis (for ON/O

Page 57

11Fig. 11.C RELAY OUTPUTS 15 to 19 WIRINGThe outputs from OUT 10 to 19 are equipped with relays havingcontact rating equal to 0.5A/250V AC on resist

Page 58

146Output limiters.For the main and/or secondary control outputs it is possible toset:- output high limits- output low limits- output max. rate of ris

Page 59

147Out 1 and 2 - SSRType: not isolated voltage output- Logic level 1:14V @ 20 mA max.24 V @ 1 mA.- Logic level 0:< 0.5 V D.C.Out 1 and 2 - TRIACSwi

Page 60

148Output type: Isolated output programmable as- 0-20 mA- 4-20 mA.Scaling: programmable from -1999 to 9999.Maximum load: 600 W.Accuracy:- 0.1 % whe

Page 61

149Deviation alarmOperative mode: High or low programmable.Threshold: programmable from - 1000 to +1000 digits.Hysteresis: programmable in engineering

Page 62

150AUXILIARY POWER SUPPLYType: 24 V DC + 20% not isolated.Max. current: 25 mA.SERIAL INTERFACETypes: RS 485 isolatedProtocol type: MODBUS, JBUS.Baud r

Page 63

A. 1DEFAULT PARAMETERSGeneral notes:A complete and consistent set of run time and configurationparameters is memorized in the instrument. These data a

Page 64

A. 2B.4) Push the FUNC pushbutton.The middle display will show:and then it will show:The default parameter loading procedure of all run timeparameters

Page 65

A. 3Run time group 4 [r.dxx]PARAMETER DEFAULT VALUEPb = 4.0 %HYS = 0.5 %ti = 4.00 mm.sstd = 1.00 mm.ssIP = 50.0 (If only one control output is configu

Page 66

A. 4Pb 5 = 4.0 %HYS 5 = 0.5 %ti 5 = 4.00 mm.sstd 5 = 1.00 mm.ssIP 5 = 50.0 (If only one control output is configured).0.0 (If two control outputs, wit

Page 67

A. 5A3.Ac = rEVA3.St = OFFA4.tP = ProcA4.Cn = H.A.A4.Ac = rEVA4.St = OFFFd.Cn = A.Fd.Ac = rEVRun time group 7 [r.Gxx]PARAMETER DEFAULT VALUES.L.Pr = ñ

Page 68

12The same problem may occur when a switch is used in serieswith the internal contacts as shown in Fig. 12.Fig. 12 EXTERNAL SWITCH IN SERIES WITH THE

Page 69

A. 6Run time group Hd [r.Lxx]PARAMETER DEFAULT VALUEPb.Lo = 1.0%Pb.Hi = 300.0%ti.Lo = 00.20 mm.ssti.Hi = 20.00 mm.ssrG.CL = OFFDEFAULT CONFIGURATION A

Page 70

A. 7- turned OFF the bargraph displays (MKC only);- sets analog retransmissions to the retransmitted initialscale value;- sets alarms in no alarm cond

Page 71 - Preliminary

A. 8A.In.F nonEA.In.t 4-20 mAA.In.L 0 °CA.In.H 1000 °CA.I.FL 0 (No filter)Configuration group 2 [C.Exx]PARAM. VALUE NOTESO1.Fn ñAin If servo motor dr

Page 72 - CONTROL PARAMETERS

A. 9SC.dP ----. (No decimal figure)SC.E.L 0SC.E.H 100SC.A.C bEFrConfiguration group 4 [C.Gxx]PARAM. VALUE NOTESSñ.Fn EnbCn.tP PIdñAn.F EnbAñ.UL buñ.ñ.

Page 73

A. 10F.E.10 nonES.E.10 OPEnF.E.11 nonES.E.11 OPEnF.E.12 nonES.E.12 OPEnF.E.13 nonES.E.13 OPEnF.E.14 nonES.E.14 OPEnF.E.15 nonES.E.15 OPEnF.E.16 nonES.

Page 74

A. 11TABLE 2Configuration group 1 [C.dxx]PARAM. VALUE NOTESLn.Fr 60 Hzñ.In.t 18 (TC K with °F)ñ.In.d----. (No decimal figure)ñ.In.S dIS Disabledñ.In.L

Page 75

A. 12Configuration group 3 [C.Fxx]PARAM. VALUE NOTESSPLt. dISñC.Gn 2.00ñC.bS -50.0SC.Gn 2.00SC.bS 0.0ñC.Cn norññ.SCL nOñC.dP----. (No decimal figure)ñ

Page 76

A. 13I2.Fn nonEI2.St OPEnI3.Fn nonEI3.St OPEnI4.Fn nonEI4.St OPEnI5.Fn nonEI5.St OPEnI6.Fn nonEI6.St OPEnI7.Fn nonEI7.St OPEnI8.Fn nonEI8.St OPEnF.E.0

Page 77

Configuration group 6 [C.Ixx]PARAM. VALUE NOTESG.brG DEvO.brG P.OutbrG.L -150 °FbrG.H 2500 °FbrG.d 10 °FSP.AL n.ALGSP.dS OP.SPSr.bH bALSS.tr -150 °Ft.

Page 78

XkP-A-C2.P65 10/30/01, 9:44 AM15

Page 79

13Switching mode: isolated zero crossing type.Rated current: from 50 mA to 1 A.Rated voltage: from 24 VRMS to 240 VRMS -10 % +15 % (50/60Hz)Load type

Page 80

Cincinnati, OH800-666-4330http://www.supersystems.comManual Part No. 31090 20QCopyright © 2001 Super Systems Inc.XkP-A-C2.P65 10/30/01, 9:44 AM16

Page 81

14The two relay output must be interlocked (see chapter"Preliminary hardware setting" paragraph "Out 3 and 4selection").NOTES:1) B

Page 82

15E.5) ANALOG OUTPUTSFig. 16.A OUTPUT 5 WIRINGFig. 16.B OUTPUT 6 WIRINGNOTE:1) Do not run analog output wirings together with AC powercables.2) O

Page 83

XkP-0-1C2.P65 10/30/01, 9:42 AM2

Page 84

163) The EIA standard establishes that by RS-485 interface it ispossible to connect up to 30 devices with one remote masterunit.The serial interface o

Page 85

176) For 24 V DC the polarity is a not care condition.7) The power supply input is fuse protected by a sub miniature fuserated T, 1A, 250 V.When fuse

Page 86

18PRELIMINARY HARDWARE SETTINGSHow to remove the instrument from its case1) Switch off the instrument.2) Push gently the lock A on the right.3) While

Page 87

19 1 3 5 7 2 4 6 8 1 3 5 7 2 4 6 8J102J103CPUcard1 2 3 4ON DIPV301J205MAIN INPUT SELECTIONSet

Page 88

20Output J 204 J 205Relay close openServo open closeNOTE: when the servomotor close loop or the servomotor openloop with valve position indication out

Page 89

21Fig. 22To select the desired input type, set V301 (see fig. 20) asdetailed in the following table:Input V301.1 V301.2 V301.3 V301.4IN CT ON OFF ON O

Page 90

22Operative mode and Hardware lockBy V101 (see fig 22) it is possible to select one of the followingoperative modes:a) run time mode without configura

Page 91

23Security code setting:1) Remove the instrument from its case.2) Set the internal dip switch V101 as follows:- V101.1 = OFF- V101.2 = ON- V101.3 = OF

Page 92

24Configuration security codeThe display will show:Note: the middle display shows the current status of theconfiguration security code ("0",

Page 93

25NOTE: at the end of the security code setting procedure, setV101 according to the desired operative mode (see"Operative mode and hardware lock&

Page 94

INDEXMOUNTING REQUIREMENTS ... 1OUTLINE AND CUT OUT DIMENSIONS ... 2CONNECTION

Page 95

26Keyboard descriptionMENU = is used to select a parameter groupFUNC = o when the instrument is in "normal display mode" itchanges the indic

Page 96

27t+MENU= are used to start the lamp test function (seeparagraph "Lamp test")s+FUNC or t+FUNCDuring parameter modification they allow you to

Page 97

28By s or t push-button it is possible to select between: = (monitor) this selection allows you to monitor butnot to modify the value assigned to the

Page 98

29CnF. 12) During monitor mode, if no push-button is depressed for morethan 10 s (or 30 s according to "t.out" [C.I10] parametersetting), th

Page 99

30CnF. 1specific instrument configuration (i.e. setting OUT 3 differentfrom servo, all the parameters related to servomotor outputwill be skipped).2)

Page 100

31CnF. 18 = TC S From - 50 to 1760 °C9 = TC R From - 50 to 1760 °C10 = TC B From 0 to 1820 °C11 = TC G (or W) From 0 to 2300 °C12 = TC D (

Page 101

32CnF. 12) When the square root extraction is enabled the values of thefollowing parameters:- "ñ.In.L" (read-out - initial scale value [C.d0

Page 102

33CnF. 1Real curveRead-outAdjustedcurveInputOFSt2) Changing the value of this parameter, the "brG.L" (bargraph -initial scale value [C.I03])

Page 103

34CnF. 1- Auxiliary input function - [C.d09]Range: nonE = Input not usedbIAS = Input used as bias for local set pointNOTES:1) When auxiliary input opt

Page 104 - PROGRAMMER MODE

35CnF. 2- Filter on auxiliary input - [C.d13]This parameter is available only when auxiliary input isconfiguredRanges: from 0 (no filter) to 8 seconds

Page 105

XkP-0-1C2.P65 10/30/01, 9:42 AM4

Page 106

36CnF. 2- OUT 2 function - [C.E02]Range: nonE = Output not usedñAin = Time proportional main control outputSECn = Time proportional secondary control

Page 107

37CnF. 2- Valve position indication - [C.E06]This parameter will be displayed only when open loopservomotor control drive output is configured.Range:

Page 108

38CnF. 2- OUT 5 retransmission full scale value- [C.E10]This parameter will be available only when Out 5 function[C.E07] is configured as “PV.rt” or “

Page 109

39CnF. 2- OUT 6 retransmission full scale value- [C.E15]This parameter will be available only when Out 6 function[C.E12] is configured as “PV.rt” or “

Page 110

40CnF. 2II) Exiting from the configuration group 2 also the followingactions are automatically performed:A) The "Añ.UL” parameter ("Output

Page 111

41CnF. 3resulting real outputs are shown below:where:- for the first split output (MAIN)Bias 1 = -AGain 1 = 100 / (B - A)- For the second split output

Page 112

42CnF. 3We will set:Bias 1 = 0Gain 1 = 100 / (33.3 - 0) = 3Bias 2 = - 33.3Gain 2 = 100 / (100 - 33.3) = 1,5The bias and gain of the two split outputs

Page 113

43CnF. 3- Main control output conditioning - [C.F06]This parameter will be available only when main control output isconfigured.Range: norñ = The cont

Page 114

44CnF. 3- Main output initial scale readout - [C.F09]This parameter will be available only when "ñ.SCL" ("Maincontrol output in enginee

Page 115

45CnF. 3- Main output auxiliary conditioning - [C.F11]This parameter will be available only when main control output isconfigured and “ñC.Cn” ("M

Page 116

1MOUNTING REQUIREMENTSThis instrument is intended for permanent installation, for indooruse only, in an electrical panel which encloses the rear housi

Page 117

46CnF. 3- Secondary control output in engineering unit- [C.F13]This parameter will be available only when secondary controloutput is configured.Range:

Page 118

47CnF. 3- Secondary output auxiliary conditioning- [C.F17]This parameter will be available only when secondary controloutput is configured and “SC.Cn”

Page 119

48CnF. 4CONFIGURATION GROUP 4 [C.Gxx]AUXILIARY CONTROL CONFIGURATION- Smart function - [C.G01]This parameter will be available only when at least one

Page 120

49CnF. 4NOTES:1) When open loop servomotor control drive without valveposition indication is configured, this parameter is forced to“buñ.” and it cann

Page 121

50CnF. 4- Program restarting after a power failure- [C.G07]Range: Edit = the program execution will be stopped and theinstrument starts in STAND-BY mo

Page 122

51CnF. 5- Program restart tracking band - [C.G08]This parameter will be available only when "St.Pr" (programrestart after power supply failu

Page 123

52CnF. 5CONFIGURATION GROUP 5 [C.Hxx]DIGITAL INPUT/OUTPUT CONFIGURATION- Logic input "DIG 1" function - [C.H01]This parameter will be availa

Page 124

53CnF. 5Notes about logic inputs used for program selection1) When some logic inputs are programmed as programselectors ("Pr.SL") the instru

Page 125

54CnF. 5NOTE: when an input is programmed as RUN/HOLD selector related with the status, the RUN/HOLD selection by keyboard has no effect.ru.St = Input

Page 126

55CnF. 5NOTE: See also all NOTES of the "d1.Fn" ("Logic input 1function" [C.H01]) parameter.- Logic input DIG 2 - contact status -

Page 127

2OUTLINE AND CUT OUTDIMENSIONSFig. 2.B OUTLINE AND CUT-OUT DIMENSIONSXKP-1-C2.p65 10/30/01, 9:36 AM2

Page 128

56CnF. 5Au.ñA = Input contact used for Auto/Manual selection(Manual when logic level is “1”)O.LIñ = Input contact used for output limiter activation(O

Page 129

57CnF. 5NOTE: for ABORT function only, the inputstatus must be maintained for more than 3seconds.Pr.SL = Input contact used for program selection (for

Page 130

58CnF. 5the output programmed as "end of profileannunciator" in ON status for a time equal tothe value assigned to "tE.Pr" [C.H12]

Page 131

59CnF. 6- Time for the "End of cycle" annunciator- [C.H11]Available only when one of the Events is programmed as "Endof cycle" ann

Page 132

60CnF. 6CONFIGURATION GROUP 6 [C.Ixx]OTHER CONFIGURATION PARAMETERS- Green bargraph function - [C.I01]This parameter will be available for 25Q models

Page 133

61CnF. 6- Resolution of the deviation bargraph - [C.I05]This parameter will be available only for 25Q models when“G.brG” [C.I01] is equal to “DEV.”.Ra

Page 134 - CONTOLLER MODE

62CnF. 6- Servomotor behaviour when the outputpower is limited - [C.I08]This parameter will be available only when the close loopservomotor control dr

Page 135

63CnF. 6- Out failure detection (OFD) - [C.I11]This parameter will be available only when the option is fittedand the main and/or the secondary output

Page 136

64CnF. 6- Automatic start enabling - [C.I14]This parameter will be available only when the clock calendarhardware is mounted.Range: nonE = The "A

Page 137

65VErFCONFIGURATION GROUP VERF [C.Lxx]CLOCK AND INPUT/OUTPUT TESTThis group is displayed only when configuration modify mode isenabled.This group allo

Page 138 - SMART FUNCTION

3Connections are to be made with the instrumenthousing installed in its proper location.A) MEASURING INPUTSNOTE: Any external component (like zener ba

Page 139

66VErF- Clock calendar test - [C.L01]This parameter is available only if the clock calendar option isfitted.Middle display: minutes and secondsNOTE: T

Page 140

67General note for configuration group "VErF"Returning in the "run time mode", the instrument restarts tocontrol the process, the

Page 141

68ATTENTION: One software key protects one or more run timeparameter groups so that when the software key is set, it ispossible to modify the run time

Page 142

69NOTES:1) In the following pages we will describe all the parameter ofthe instrument. The instrument, however, will show only theparameters related

Page 143

70Gr. 2Run time group 2 [r.bxx]COMMAND- Programmer/controller mode selection- [r.b01]This parameter can be modified only when no program isrunning.Ran

Page 144

71Gr. 3from 1 to 90 = it is the simple program selected for theclock start.from L1 to L9= It is the linked program selected for theclock start.NOTES:1

Page 145 - GENERAL INFORMATIONS

72Gr. 3tHU = ThursdayFri = FridaySat = SaturdaySUn = SundayNOTE: this is a readable parameter only and it cannot bemodified.- Group 2 default data loa

Page 146

73Gr. 3Range: - in engineering units within the span limits for processalarm;- from -1000 to 1000 digits for deviation alarm;- from 0.0% to 100.0 % f

Page 147

74Gr. 3- high threshold used when the alarm 2 is aband alarm - [r.C07]This parameter is available only when the alarm 2 is configuredas band alarm (“A

Page 148

75Gr. 3NOTE: The span limits are configured by “ñ.In.L” [C.d05] and“ñ.In.H” [C.d06].- Low threshold used when the alarm 4 is aband alarm - [r.C12]This

Page 149

4A.1) TC INPUTFig. 4 THERMOCOUPLE INPUT WIRINGNOTES:1) Don’t run input wires together with power cables.2) For TC wiring use proper compensating cabl

Page 150

76Gr. 3- Alarm 1 hysteresis - [r.C16]This parameter is available only when Out 1 is configured asalarm output (“O1.Fn” [C.E01] = “ALr.1”)Range: from 1

Page 151

77Gr. 4Run time group 4 [r.dxx]CONTROL PARAMETERSNOTE: when SMART function is activated, all the Group 4parameters cannot be modified.- Proportional b

Page 152

78Gr. 4NOTES:1) When device is working with SMART algorithm and “Cn.tP”[C.G02] is equal to “Pid”, the “td” value will be made equal toa quarter of “t

Page 153

79Gr. 4NOTES:1) When device is configured with two control outputs (splitrange excluded) the control action is forced to reverse and itcannot be modif

Page 154

80Gr. 4- Hysteresis (for ON/OFF control) of the firstPID group - [r.d10]This parameter is available when Pb [r.d01]=0 (On/OFF controlaction) and the i

Page 155

81Gr. 5- Integral preload of the first PID group- [r.d13]This parameter is available when at least one control output isconfigured, "Pb" [r.

Page 156

82Gr. 5Run time group 5 [r.Exx]AUXILIARY CONTROL PARAMETER- Anti reset wind up - [r.E01]This parameter is available when at least one control output i

Page 157

83Gr. 5- Main control output max rate of rise - [r.E06]This parameter is available when one control output isconfigured as main output.Ranges: from 0.

Page 158

84Gr. 5- Set point high limit - [r.E13]Range: from rL [r.E12] to “ñ.In.H” [C.d06].NOTES: Whenever “ñ.In.H” [C.d06] has been modified, thefollowing occ

Page 159

85Gr. 6- Time out for the soft start - [r.E16]This parameter is available when at least one output isconfigured as control output and no logic input i

Page 160

5Shield+_TX31117A.3) LINEAR INPUTFig. 6 mA, mV AND V INPUTS WIRINGNOTES:1) Don’t run input wires together with power cables.2) Pay attention to the

Page 161 - ----. (No decimal figure)

86Gr. 6Run time group 6 [r.Fxx]ALARM SETTING- Alarm 1 type - [r.F01]This parameter is available only when OUT 1 is configured asalarm 1 output (“O1.Fn

Page 162

87Gr. 6Example for A1.Cn (A2.Cn, A3.Cn) = H.A.A= alarm condition detectionB= automatic reset of alarmC= manual reset; the LED gets steady lit, the ala

Page 163

88Gr. 6Example for A1.Cn (A2.Cn, A3.Cn) = H.L.A= alarm condition detectionB= the alarm remains in alarm status (even if the measuredvalue is under the

Page 164

89Gr. 6NOTES:1) If the alarm is programmed as band or deviation alarm, thisfunction masks the alarm condition after a set point change orat the instru

Page 165

90Gr. 6- Alarm 2 action - [r.F07]This parameter is available only when OUT 2 is configured asalarm 2 output (“O2.Fn” [C.E02] = “ALr.2”)Range: dir = di

Page 166

91Gr. 6- Alarm 3 configuration - [r.F10]This parameter is available only when OUT 3 is configured asalarm 3 output (“O3.Fn” [C.E03] = “ALr.3”)Range: H

Page 167

92Gr. 6- Alarm 4 type - [r.F13]This parameter is available only when OUT 4 is configured asalarm 4 output (“O4.Fn” [C.E04] = “ALr.4”)Range: Proc = Ala

Page 168

93Gr. 6alarm threshold plus or minus hysteresis. If the alarm isprogrammed as a process alarm or control output alarm, thisfunction masks the alarm co

Page 169

94Gr. 7Run time group 7 [r.Gxx]SERIAL LINK PARAMETER- Serial interface protocol - [r.G01]Range: OFF = No serial interfaceñbUS = ModbusjbUS = JbusNOTE

Page 170 - Manual Part No. 31090 20Q

95Gr. 8Run time group 8 [r.Hxx]FEEDBACK POTENTIOMETER ADJUSTMENTNOTE: The run time group 8 will be displayed only if theinstrument is in MANUAL mode.-

Commentaires sur ces manuels

Pas de commentaire