X5 Controller
Calibration Guide
Super Systems Inc. X5 Controller: Calibration Guide. (Document #4615; New: 05/01/2014).
Visit us online at http://www.supersystems.com.
Introduction
The instrument calibration is based on a series of steps. Each step performs the calibration of a particular input or
output. It must be done only when strictly necessary and using reference instruments with a high accuracy. Then the
instrument must remain switched on at current environmental conditions for at least one hour.
Calibration Procedure
Starting from operation mode, press and hold both the “ “(enter) and “(menu access) keys. C.PaSwill
appear. Enter the Configuration Password (default: 33).
NOTE: See the X5 Controller product manual section “4.6 PARAMETERIZATION - ACCESS MENU - PASSWORD
- CALIBRATIONfor more details on the Configuration Password.
C.tc” (T/C calibration) will appear on the lower display while a value will appear on the upper display.
Generate a signal using the below table;
use copper wire to source the signal. Press the “ “ or ““ keys to
adjust the value to be as close to zero as possible: 0 ±2.
Step
Function
Action
C.tc.
TC calibration
Apply 50.000mV to the input set with
or
keys
C.Pt.
RTD calibration
Apply 313.594 to the input set with or keys
C.mV.
mV calibration
Apply 300 mV to the input set with or keys
C.Vol
Volts Calibration
Apply 10 V to the input set with
or
keys
C.Ma
mA calibration
Apply 20mA to the input set with or keys
C.S. V
Remote Setpoint Volts
calibration
Apply 10 V to the input set with
or
keys
C.S. mA
Remote Setpoint mA
calibration
Apply 20mA to the input set with
or
keys
Table 1 - Calibration types and actions
Wait until the displayed value remains stable: 0 ± (2 to 5).
IMPORTANT: Press the “” key to bypass a calibration step. DO NOT press the “ “ or “keys unless you
want to Calibrate that specific step.
When the calibration is correct, press the “” key to go to the next step. It is possible to not modify the
calibration. When the step has been displayed on screen DO NOT press the “
“ or “ “ keys; only press the
” key until out of calibration mode.
–OR-
After “C.tc.”, press the “” key to re-enter normal operation.